Attaching Accessories to the Operating Table Top
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
59
15
Attaching Accessories to the Operating Table Top
A typical accessory is the radial adjustment clamp with the Goepel-
type leg support. The upper cleat of the single clamp is fixed.
Attaching the
accessory
1. Release the locking handle from the clamp. The clamp opens.
2. Hook the accessory part [A] with the claw obliquely from
above into the side rail.
3. Turn the locking handle on the clamp to tighten. The clamp
closes.
CAUTION!
Risk of personal injury. Risk of injury due to improper
attachment!
Unlocked accessories can fall away from the operating table top
and cause injury to staff or patients. Observe the following points if
other parts are attached to the operating table top:
– Follow the user manual of the accessories.
– Check that the accessory is securely attached to the operating
table top before use.
– Never attach or detach several accessories simultaneously.