Load Capacity
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
67
18.1.3
Operating table top with seat section extension and upper back section
[28] Head section
[30] Seat section extension
[31] Upper back section
1. No section segment is necessary at the head end of the
operating table top.
2. A seat section extension [30], upper back section [31] and a
head section [28] are attached at the foot end of the
operating table top.
3. Patient orientation on the operating table top is inverse.
Permissible load:
Attention, additional accessories reduce the permissible patient
weight.
18.1.4
Operating table top with Carbon 600 table top segment
[32] Carbon 600 table top segment
[33] X-RAY head positioning accessory
1. No section segment is necessary at the head end of the
operating table top.
2. A Carbon 600 table top segment [32] with X-RAY head-
positioning accessory [33] is fastened to the foot end of the
operating table top.
3. Patient orientation on the operating table top is inverse.
Permissible load:
Attention, additional accessories reduce the permissible patient
weight.
496 lbs
(225 kg)
A patient weight of up to 496 lbs (225 kg) is possible
with restricted longitudinal travel toward the head end
of the patient.
353 lbs
(160 kg)
A patient weight of up to 353 lbs (160 kg) is possible
without limiting the operating table functions.
353 lbs
(160 kg)
A patient weight of up to 353 lbs (160 kg) is possible
with restricted longitudinal travel toward the head end
of the patient.