Emergency Mode
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
93
22
Emergency Mode
If there is a defect in the central operating table control, the
emergency mode enables ongoing functions to continue until
complete and the patient to be transferred.
Restriction
During emergency mode the operating table can only be used with
limited functions
– The operating table can only be controlled using the column
keypad.
– Functions move at a lower speed and a continuous warning tone
sounds.
– Leveling is limited. Only the column functions are leveled in
conjunction with the level position function. The table top has to
be moved to level position manually for the individual functions.
– The running gear functionality is disabled.
The emergency mode must only be activated in case of a serious
malfunction of the operating table! Please contact
TRUMPF Medical Systems, Inc., Technical Customer Services
immediately for the necessary repair! The defective operating table
must be barred from further operations!
Activating
emergency mode
The emergency mode must be manually activated by the user:
1. Unlock the column keypad. Press key [i53] for this purpose.
2. Press key [i20] on the column keypad for at least 10 seconds.
A warning tone that continually repeats is heard.
The emergency mode is acknowledged with a double tone
and then activated. Indicator [i20] lights up red continuously
and indicator [i1] flashes green.
The operating table can be set to normal mode by switching the
system off and then back on again.