Load Capacity
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
71
18.2.2
Operating table top with head section, seat section extension and leg sections
[28] Head section
[29] Leg sections
[30] Seat section extension
1. A head section [28] is attached at the head end of the
operating table top.
2. A seat section extension [30] and leg sections [29] are
attached at the foot end of the operating table top. The
permissible patient weight depends on the leg section used.
3. Patient orientation on the operating table top is as normal.
Permissible load:
Attention, additional accessories reduce the permissible patient
weight.
18.2.3
Operating table top with Carbon 600 table top segment and leg sections
[29] Leg sections
[32] Carbon 600 table top segment
[33] X-RAY head positioning accessory
1. Leg sections [29] are attached at the head end of the
operating table top.
2. A Carbon 600 table top segment [32] with X-RAY head-
positioning accessory [33] is fastened to the foot end of the
operating table top.
3. Patient orientation on the operating table top is inverse.
Permissible load:
496 lbs
(225 kg)
Patient weights of up to 496 lbs (225 kg) are possible,
provided a four-part leg section is not attached to the
operating table top. The operating table functions can
be used without limitation.
441 lbs
(200 kg)
Patient weights of up to 441 lbs (200 kg) are possible
with any leg section. The operating table functions can
be used without limitation.
353 lbs
(160 kg)
A patient weight of up to 353 lbs (160 kg) is possible
without limiting the operating table functions. Attention,
additional accessories reduce the permissible patient
weight.