Control Units
38
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
6.1.2
Optical indicators
Indicator
State
Color Meaning
Action
Ready [i1]
illuminated green Operating table is switched on
and ready.
–
flashes
green In addition to indicator [i1], the
emergency mode indicator is
illuminated.
See indicator [i20]
not
illuminated
Operating table is switched off Switch on the operating table if
necessary.
Key release [i53]
illuminated green Column keypad keys released.
The keypad locks 10 seconds
after the last key is pressed.
–
flashes
green In addition, a shrill single tone
(warning tone) sounds.
The column keypad is locked.
A locked key was pressed on the
keypad. Direct operation of the
operating table using the column
keypad is not possible.
Press button [i53]. The keypad
is released for operation for
10 seconds.
Select functions on the column
keypad using the two-key
control.
not
illuminated
The column keypad is locked.
Direct operation of the operating
table using the column keypad is
not possible.
Except for
O
FF
key
The operating table can be
turned off at any time using the
O
FF
button on the column
keypad.
Press button [i53]. The keypad
is released for operation for
10 seconds and the indicator is
constantly illuminated.
Select functions on the column
keypad using the two-key
control.
Head position left [i3]
illuminated green Indicator corresponds to the
head position of the patient on
the operating table.
–