Controlling the Operating Table
84
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
19.9
Back section joints
On a two-part table top, the back section [4] is permanently
attached to the seat section [3] using the back section joints [7]. The
back section can be tilted up or down.
On a one-part table top, the back section joints [7] can only be
tilted up or down together.
In the inverse patient position, the back section joints [7] are
controlled by the keys for the functions of the leg sections.
Limitations
The adjustment range of the joints is automatically restricted or
locked in the following situations.
1. Depending on the longitudinal travel, the downward
adjustment range of the joints may be restricted. If a larger
adjustment range is required, the longitudinal travel must be
changed.
2. The adjustment ranges for the normal and inverse patient
positions are different (see section 24 on page 96).
Adjusting the
function
Press key [i14] or [i15].
[i14] Back section joint up
[i15] Back section joint down