Controlling the Operating Table
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
75
19
Controlling the Operating Table
The operating table is operated with the remote control or via the
column keypad. For this reason, the instruction manual of the remote
control must also be followed. In the following sections, the remote
control keys of the TruSystem 7000 are shown. The symbols on the
keys may deviate in other operating units produced by
Trumpf Medical. Keys with the same function are also designated
in the same way in the user manual of the various operating units (for
example, [i9] for left tilt key).
Adjusting the
operating table
– Establish a potential connection to the running gear.
– Switch on and brake (jack up) the operating table.
– Cover operating table with absorbent towels.
– Transfer the patient onto the operating table.
– Compare the position of the head with the indication on the
column keypad and alter if necessary.
– Secure the patient on the operating table.
– Adjust the functions on the operating table.
Adjusting the
function
Hold down the function key on an operating unit until the desired
position is reached. Function stops in the following situations:
– The key is released.
– the level position has been reached.
– an intermediate stop has been reached (see section 20 on
page 86).
– The end position is reached.
The automatic stop and the end position are indicated by an
audible signal. For any further adjustment, briefly release the
function key and then press it again.
CAUTION!
Risk of personal injury. Hazard to the patient due to risk of
collision! Risk of entrapment on the part of the operator!
The operating table does not recognize objects within the
surrounding environment. Collisions with the furnishings or the
devices that are located below the operating table top must be
prevented by the operator. In addition, with complete operating
table equipment, not all equipment parts are recognized by the
operating table electronic system. Collisions therefore cannot be
completely excluded. Observe all motorized movements of the
operating table until the end position and stop the function before
a hazardous situation occurs. Clear the area beneath the
operating table top.