Controlling the Operating Table
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
79
19.4
Tilt
This function tilts the operating table top to the right or left about its
longitudinal axis.
Limitations
The adjustment range of the tilt is automatically restricted in the
following situations.
1. The adjustment area is limited to +10°/–10° in drive mode.
2. The tilt may be restricted depending on the (reverse)
Trendelenburg position The adjustment range depends on the
patient's weight. For heavy patients, the adjustment range is
smaller. The (reverse) Trendelenburg position needs to be
reduced if a larger setting range is required.
Adjusting the
function
Press key [i9] or [i10]. For safety reasons, the tilt function stops
automatically at maximum 15° (see section 20 page 86).
19.5
Anti (Reverse)-Trendelenburg/Trendelenburg
The Anti-Trendelenburg/Trendelenburg function is used to incline
the operating table top around its transverse axis toward the foot
end or head end.
Limitations
Depending on the tilt, the (reverse) Trendelenburg position may be
automatically restricted. The adjustment range depends on the
patient's weight. For heavy patients, the adjustment range is smaller.
If a larger adjustment range is required, the tilt must be reduced.
[i9] Tilt left
[i10] Tilt right