Orion Benchtop pH/ISE Meter Instruction Manual
Overview of Meter Functions 520Aplus,525Aplus,720Aplus and 920Aplus 109
Automatic Datalog Function
The meters have an internal datalogger that will record and store a
maximum of 25 points automatically. The data points remain in memory
until cleared.
Prompt
Keypress/Action
Press
data/print
modes
1. MANUAL PRINT
The current print mode is displayed. Wait
for the prompt.
2. SET PRINT MODE ?
Press
no
.
3. DATALOG
The current datalog setting is displayed.
MANUAL
Wait for prompt.
4. SET LOG MODE
Press
yes
.
5. 0-DISABLE 1-RDY
The datalog setting options are presented
2-TIM 3-MAN
two at a time. Press
no
to display the other
two. Enter
0
,
1
,
2
or
3
, then press
yes.
0 To disable datalogging.
1 For datalogging on ready.
2 For datalogging at present time
interval.
3 For manual datalogging when print
is pressed.
6. DATALOG
The meter verifies the setting and returns to
MANUAL
measure mode.
7. TIMED LOG
If
2
(datalog at a preset time interval) is
selected the interval must be entered at the
following prompts.
8. DATALOG INT HR
Enter the datalog interval hours setting, from
0 to 23, then press
yes
.
9. DATALOG INT MIN
Enter the datalog interval minutes setting,
from 0 to 59, then press
yes
.
10. DATALOG INT SEC
Enter the datalog interval seconds setting
from 0 to 59, then press
yes
. Logging can
not take place faster than every 5 seconds.
11. DATALOG
The datalog interval is displayed then the
INTERVAL
meter returns to MEASURE mode.
Summary of Contents for Orion 410Aplus
Page 6: ...2 Orion Benchtop pH ISE Meter Instruction Manual...
Page 84: ...80 Orion Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Page 92: ...Orion Benchtop pH ISE Meter Instruction Manual 88...
Page 104: ...Orion Benchtop pH ISE Meter Instruction Manual 100...
Page 120: ...Orion Benchtop pH ISE Meter Instruction Manual 116...
Page 122: ...Orion Benchtop pH ISE Meter Instruction Manual 118...
Page 126: ...Orion Benchtop pH ISE Meter Instruction Manual 122...