C. Orion 420Aplus and 710Aplus Check-Out Procedure
1. Perform the self-test.
2. After completing the self-test the meter will be in MEASURE mode.
The legend MEASURE will be displayed.
a. Press
mode
until the pH mode indicator is displayed. Main display
should read a steady 7.00 ± 0.02.
NOTE
: If this is the first time this procedure has been
performed the reading should be 7.00 ± 0.02.
b. If not, press
2nd cal
. “P1” will appear. When “Ready” appears,
press
yes
.
c. Press
measure
. The main display should read 100.0 with the
legend SLP in the lower display. If so, press
yes
.
d. If not, scroll until the display reads 100.0 and then press
yes
.
The meter advances to MEASURE and the display should
now read steady 7.00.
NOTE
: To change a value, press or . The first digit
will flash, continue scrolling until the first digit equals the
correct value, then press
yes.
The second digit will flash.
Scroll to the correct value then press
yes
. When all digits
have been changed, press
yes
to enter the new value.
3. Press
mode
to enter millivolt mode. Display should read 0.0 ± 0.2. If
not, press
2nd cal
then press
yes
to enter the value 0.0. The meter will
return to MEASURE mode.
4. Press
mode
to enter REL mV mode. Display should read 0.0 ± 0.2. If
not, press
2nd cal
then press
yes
to enter the value 0.0. The meter will
return to MEASURE mode.
Orion Benchtop pH/ISE Meter Instruction Manual
Set Up, Self-Test and Check-Out
6
▼
▼
Summary of Contents for Orion 410Aplus
Page 6: ...2 Orion Benchtop pH ISE Meter Instruction Manual...
Page 84: ...80 Orion Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Page 92: ...Orion Benchtop pH ISE Meter Instruction Manual 88...
Page 104: ...Orion Benchtop pH ISE Meter Instruction Manual 100...
Page 120: ...Orion Benchtop pH ISE Meter Instruction Manual 116...
Page 122: ...Orion Benchtop pH ISE Meter Instruction Manual 118...
Page 126: ...Orion Benchtop pH ISE Meter Instruction Manual 122...