Orion Benchtop pH/ISE Meter Instruction Manual
Overview of Meter Functions-420Aplus, 710Aplus
98
Code
Function
Description
5-2
SET PRINT
Sets the timed print interval. The maximum
print INTERVAL is 23 hours, 59 minutes,
and 59 seconds. The minimum print interval
is 5 seconds. When the code 5-2 is displayed
in the lower field, the current print interval
hours will flash in the main display.
Press
yes
or press or to change, then
press
yes
. For the print interval minutes:
seconds will be displayed in the main field.
Press
yes
or change, then press
yes
. The
default setting is 1 minute.
The following set-up modes apply to the 710Aplus model only:
6-1
DATA-
Sets the datalogger mode. Up to 25 points
LOGGERS
may be stored in mode memory at any one
time. When the code 6-1 is displayed in the
lower field the current datalogger mode will
flash in the main display. Use or to
change, then press
yes
.
The four options are:
0
Off, when 0 is selected no
datalogging of any kind will take place.
1
Manual datalog, no automatic output
to the datalogger will occur, the user
may datalog on command by pressing
2nd
then
.
2
Datalog on READY, datalogging occurs
when the electrode signal is stable.
3
Datalog on a timed interval. Datalogging
occurs at user selected intervals. The
default setting is 0: OFF.
▼
▼
▼
▼
Summary of Contents for Orion 410Aplus
Page 6: ...2 Orion Benchtop pH ISE Meter Instruction Manual...
Page 84: ...80 Orion Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Page 92: ...Orion Benchtop pH ISE Meter Instruction Manual 88...
Page 104: ...Orion Benchtop pH ISE Meter Instruction Manual 100...
Page 120: ...Orion Benchtop pH ISE Meter Instruction Manual 116...
Page 122: ...Orion Benchtop pH ISE Meter Instruction Manual 118...
Page 126: ...Orion Benchtop pH ISE Meter Instruction Manual 122...