pH Calibration Procedure
1. Press
2nd
then
channel
until the correct input channel is selected.
2. Press
mode
until the pH mode indicator is displayed.
3. Press
calibrate
. CALIBRATE is displayed. The time and date of the
last calibration are displayed.
4. After a few seconds ENTER NO. BUFFERS is displayed. Enter the
number of buffers;
1
,
2
, or
3
, then press
yes
.
5. When the BUFFER 1 prompt appears, rinse electrode(s) and place
into first buffer. Wait for a stable display.
6. When the electrode signal has stabilized, the READY CAL AS
prompt will appear. Press
yes
if the value is correct for your buffer. If
performing a manual calibration, use the numeric keys to change the
value, then press
yes
.
7. If more than one buffer was selected, repeat steps 5 and 6 for
each buffer.
8a. If one buffer was selected, the current electrode slope in memory is
displayed at the SLOPE prompt. Press
yes
, or use numeric keys to
enter correct value, then press
yes
. If actual slope is unknown, enter
100.0 or perform a two buffer calibration.
8b. If two or more buffers were selected, the average electrode slope is
displayed at the SLOPE prompt.
9. The meter automatically proceeds to MEASURE mode.
10. Rinse electrode(s) and place into sample. Record the pH value when
READY is indicated or when the electrode signal is stable.
Temperature is displayed in the prompt line.
Orion Benchtop pH/ISE Meter Instruction Manual
Calibration and Measurement of pH
24
Summary of Contents for Orion 410Aplus
Page 6: ...2 Orion Benchtop pH ISE Meter Instruction Manual...
Page 84: ...80 Orion Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Page 92: ...Orion Benchtop pH ISE Meter Instruction Manual 88...
Page 104: ...Orion Benchtop pH ISE Meter Instruction Manual 100...
Page 120: ...Orion Benchtop pH ISE Meter Instruction Manual 116...
Page 122: ...Orion Benchtop pH ISE Meter Instruction Manual 118...
Page 126: ...Orion Benchtop pH ISE Meter Instruction Manual 122...