Orion Benchtop pH/ISE Meter Instruction Manual
Special Features
43
Single Known Subtraction Measurement Procedure
In known subtraction, each addition of standard subtracts an increment of
the species to be measured. The known subtraction procedure and meter
prompts are the same as the known addition procedure except that the meter
asks for the reaction ratio. At this prompt enter the correct reaction ratio of
the species to be measured with the standard to be added. For example:
A sulfide (S
=
) sample is to be subtracted by the addition of silver (Ag
+
).
Two Ag
+
ions are required to subtract each S
=
ion, therefore the reaction
ratio is 0.5.
NOTE:
Make sure the correct reaction ratio of
sample to standard is entered.
Total concentration of the sample should be approximately halved upon
addition of the standard. Therefore sample concentration should be known
within an order of magnitude.
See the following table to choose the correct standard concentration and
volume of addition.
Milliliters of standard
Standard should be
to be added per 100
X times expected sample
mL of sample
concentration
1 mL
50X
5 mL
10X
10 mL
5X
Select 2-SUBTR for the known subtraction procedure then follow the
meter prompts.
Summary of Contents for Orion 410Aplus
Page 6: ...2 Orion Benchtop pH ISE Meter Instruction Manual...
Page 84: ...80 Orion Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Page 92: ...Orion Benchtop pH ISE Meter Instruction Manual 88...
Page 104: ...Orion Benchtop pH ISE Meter Instruction Manual 100...
Page 120: ...Orion Benchtop pH ISE Meter Instruction Manual 116...
Page 122: ...Orion Benchtop pH ISE Meter Instruction Manual 118...
Page 126: ...Orion Benchtop pH ISE Meter Instruction Manual 122...