Orion Benchtop pH/ISE Meter Instruction Manual
Calibration and Measurement of Concentration
35
D. Orion 920Aplus Concentration Calibration and
Measurement
Channel Selection and Electrode Identification
1. Securely connect electrodes to either input 1 or 2. The number of the
electrode input in use appears directly before the electrode ID in the
prompt line. Example: 1-X. To select a different channel press
2nd
then
channel
.
2. Identify the electrode correctly. Press
2nd
then
electrode id
to access
the electrode ID function. The channel and current electrode
identification are displayed. If correct, press
yes
. Press
no
to
scroll through choices listed below.
1-pH 2-O
2
3-F
-
4-Cl
-
5-Br
-
6-I
-
7-NO
3
-
8-CN
-
9-NH
3
10-Cl
2
11-NO
X
12-CO
2
13-Na
+
14-K
+
15-Ag
+
16-S
=
17-Ca
++
18-Cu
++
19-Pb
++
20- Cd
++
21-ClO
4
22- BF
4
-
23- SCN
-
24-KF 25-REDOX
26-X
+
27-X
-
28-
X
++
29-
X
- -
3. Enter the number of your choice then press
yes
.
4. Meter displays new choice then returns to MEASURE mode.