Orion Benchtop pH/ISE Meter Instruction Manual
Other Measurement Modes
49
B. Orion 420Aplus, 710Aplus Millivolt Measurements
The Orion 420Aplus and 710Aplus can be used to measure absolute or
relative millivolts. The millivolt modes are useful when performing
potentiometric titrations or preparing calibration curves. Detailed
instructions for any Orion electrode are given in the electrode instruction
manual. Titration instructions are included in the Orion Redox Electrode
(Orion 9678 or 9778) Instruction Manual, or in standard analytical
chemistry texts.
Absolute Millivolts
Absolute millivolts are displayed with 0.1 mV resolution in the range of
-1600.0 to +1600.0 mV.
Access the absolute millivolt mode by pressing
mode
until the mV mode
indicator is displayed.
Relative Millivolts
Relative millivolts are displayed with 0.1 mV resolution over the range of
-1999.9 to +1999.9 mV. (Absolute millivolt range ± 1600.0).
1. Access relative millivolt mode by pressing
mode
until the REL mV
mode indicator is displayed.
2. Set the relative millivolt offset by pressing
2nd
then
cal
.
CALIBRATE and the current absolute millivolts will be displayed.
Once the signal is stable, the meter displays 0.0. Use or to enter
the desired value, or leave the setting at 0.0. Press
yes
. The meter
automatically returns to MEASURE and all relative millivolt
measurements will be based on the offset.
▼
▼
Summary of Contents for Orion 410Aplus
Page 6: ...2 Orion Benchtop pH ISE Meter Instruction Manual...
Page 84: ...80 Orion Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Page 92: ...Orion Benchtop pH ISE Meter Instruction Manual 88...
Page 104: ...Orion Benchtop pH ISE Meter Instruction Manual 100...
Page 120: ...Orion Benchtop pH ISE Meter Instruction Manual 116...
Page 122: ...Orion Benchtop pH ISE Meter Instruction Manual 118...
Page 126: ...Orion Benchtop pH ISE Meter Instruction Manual 122...