Orion Benchtop pH/ISE Meter Instruction Manual
Other Measurement Modes
57
I. Karl Fischer Titrations
A 10 microampere polarizing current is available for Karl Fischer or other
polarized electrode titrations.
Use the Karl Fischer adapter, Orion 090048.
1. Insert KF electrode pin-tip connector into the KF adapter. Connect
the other pin-tip of the electrode into the reference input.
2. Connect the BNC plug of the KF adapter to the BNC meter input.
3. Insert pin-tip connector of the KF adapter into KF input on the meter.
4. Select the mV mode and then press
measure
. On the Orion 920Aplus
identify the electrode as KF.
5. Perform titration.