Orion Benchtop pH/ISE Meter Instruction Manual
Troubleshooting
75
C. Meter Operation Problem
Table III – Meter Operation Problem
Problem
Probable Cause
Remedy
No Display
No power to meter
For Orion 410Aplus,
420Aplus, 710Aplus press
the power key. For all
models, check that
the line converter is plugged
in securely.
Autoshutoff may
When the Autoshutoff
be turned on
feature is on, the meter will
automatically shut down if
no keys have been pressed
within 10 minutes. For
Orion 410Aplus, 420Aplus
and 710Aplus press the
power key to turn the meter
on. For Orion 520Aplus,
525Aplus, 720Aplus and
920Aplus, press any
key. If this feature is not
desired, turn autoshutoff off.
Erratic Readings
Meter or electrode
Follow the meter check out
or Reading out
failure
procedure.
of Range
Check electrode operation by
following the instructions in
the electrode instruction
manual.