Orion Benchtop pH/ISE Meter Instruction Manual
Set Up, Self-Test and Check-Out
9
E. Orion 520Aplus and 525Aplus
Check-Out Procedure
1. Perform the self-test.
2.
For 525Aplus only
Press
2nd channel
until channel 1 is selected
3. After completing the self-test the meter will be in MEASURE mode.
The legend MEAS will be displayed.
NOTE:
If this is the first time this procedure has been
performed the reading should be 7.00 ± 0.02.
4. Press
1st mode
to display pH.
a. The prompt line on the display should read 25.0˚ C. If not, press
1st set temp
and enter 25.0, then press
yes
.
b. Press
1st calibrate
. Enter
1
buffer then press
yes
. Display should
read 7.00 pH. If not, enter 7.00 and press
yes
.
c. Meter should display the slope as 100.0 PCT. If not, enter 100.0
and press
yes
.
d. Meter will return to measure mode. Display should read
7.00 ± 0.02.
5. Press
mode
to display mV. Display should 0.0 ± 0.2 mV. If not, enter
0.0 and press yes.
For Orion 520Aplus the meter is now ready for use with electrodes.
Remove the shorting cap.
6.
For 525Aplus only
a. Press
2nd channel
to select input channel 2.
b. Repeat steps 4 and 5 for input channel 2.
For Orion 525Aplus the meter is now ready for use with electrodes.
Remove the shorting cap.
Summary of Contents for Orion 410Aplus
Page 6: ...2 Orion Benchtop pH ISE Meter Instruction Manual...
Page 84: ...80 Orion Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Page 92: ...Orion Benchtop pH ISE Meter Instruction Manual 88...
Page 104: ...Orion Benchtop pH ISE Meter Instruction Manual 100...
Page 120: ...Orion Benchtop pH ISE Meter Instruction Manual 116...
Page 122: ...Orion Benchtop pH ISE Meter Instruction Manual 118...
Page 126: ...Orion Benchtop pH ISE Meter Instruction Manual 122...