Orion Benchtop pH/ISE Meter Instruction Manual
Special Features
37
Chapter VII
920Aplus Special Features
A. Calibration Edit
The calibration edit feature allows you to edit a pH or concentration
calibration curve without repeating the entire procedure. This is useful in
multipoint calibrations if an entry mistake was made on one of the points.
The meter will automatically recalculate all calibration values. There are
two options in the
cal edit
function:
Remeasure
or
Edit
.
Remeasure
allows you to remeasure a particular calibration point by
placing electrode(s) in the standard (or buffer), then entering the
correct value.
Edit
has two choices: change or delete. Select change if a data entry error
was made, but electrode(s) were in the correct standard at the time of
calibration. Delete removes that point from the calibration curve.
Cal edit
presents all calibration points in memory for editing. Respond to
the prompts and edit as needed.
Cal edit
may be performed for both pH and
ISE calibrations.
If no previous calibration has been performed the meter will prompt, NO
CALIBRATION, then return to MEASURE mode. Otherwise, follow the
prompts to edit the calibration. The example provided here uses a
concentration calibration. A pH calibration would substitute the word
BUFFER for STD.
Summary of Contents for Orion 410Aplus
Page 6: ...2 Orion Benchtop pH ISE Meter Instruction Manual...
Page 84: ...80 Orion Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Page 92: ...Orion Benchtop pH ISE Meter Instruction Manual 88...
Page 104: ...Orion Benchtop pH ISE Meter Instruction Manual 100...
Page 120: ...Orion Benchtop pH ISE Meter Instruction Manual 116...
Page 122: ...Orion Benchtop pH ISE Meter Instruction Manual 118...
Page 126: ...Orion Benchtop pH ISE Meter Instruction Manual 122...