Orion Benchtop pH/ISE Meter Instruction Manual
Special Features
38
Cal Edit
Prompt
Key Press/Action
1.
Press
2nd
then
cal edit.
2. STD 1 OK?
The meter displays the current value
of standard one. Press
yes
if okay or
no
to edit.
3. If
no
selected:
Press
1
or
2
then press
yes
.
1-REMEAS 2-EDIT
4. If
1
selected:
Place electrode(s) in the standard
READY ENTER
and wait for a stable display.
VALUE
Enter the value of the standard and
press
yes.
5. If
2
selected: Press
1
or
2
then
yes
.
1-CHANGE
2-DELETE
6. If
1
selected:
The meter prompts for the new
ENTER VALUE
value. Enter value and press
yes
.
7. If
2
selected:
The meter deletes the calibration
STD 2 OK?
point and continues to the next
calibration point. After presenting all
calibration points for editing, the
meter returns to measure mode.