410Aplus
420Aplus
520Aplus
525Aplus
pH range
-2.000 to 19.999
-2.000 to 19.999
-2.000 to 19.999
-2.000 to 19.999
resolution
0.001/0.01/0.1pH
0.001/0.01/0.1pH
0.001/0.01/0.1pH
0.001/0.01/0.1pH
relative accuracy
±0.005
±0.005
±0.005
±0.005
slope
80 to 120%
80 to 120%
80 to 120%
80 to 120%
concentration
range
NA
NA
NA
NA
resolution
NA
NA
NA
NA
relative accuracy
NA
NA
NA
NA
temperature
-5.0 to 105.0oC
-5.0 to 105.0 oC
-5.0 to 105.0 oC
-5.0 to 105.0 oC
resolution
0.1 oC
0.1 oC
0.1 oC
0.1 oC
relative accuracy
±1.0 oC
±1.0 oC
±1.0 oC
±1.0 oC
millivolt range
NA
-1600.0 to +1600.0
-1600.0 to +1600.0
-1600.0 to +1600.0
resolution
NA
0.1 mV
0.1 mV
0.1 mV
relative accuracy
NA
±0.2 mV or±0.05%
±0.2 mV or ±0.05%
±0.2 mV or ±0.05%
of reading which ever is greater of reading which ever is greater
of reading whichever is greater
relative millivolt
range
NA
-1999.9 to +1999.9
-1999.9 to +1999.9
-1999.9 to +1999.9
resolution
NA
0.1 mV
0.1 mV
0.1 mV
relative accuracy
NA
±0.2 mV or ±0.05%
±0.2 mV or ±0.05%
±0.2 mV or ±0.05%
of reading whichever is greater
of readings whichever is greater
of reading whichever is greater
display
Custom LCD
Custom LCD
Custom 2-color
Custom 2-color
vacuum fluorescent
vacuum fluorescent
with17 character prompt line
with17 character prompt line
inputs
One BNC, One pin tip,
One BNC, One pin tip,
One BNC, One pin tip,
Two BNC, Two pin tip,
ATC, power, RS232
ATC, power
ATC, power, RS232, Karl Fischer
ATC, power, RS232, Karl Fischer
outputs
RS232
RS232
RS232, 2.5 mm
RS232, 2.5 mm
pin recorder
pin recorder
power
9 VDC line converter
9 VDC line converter
9 VDC line converter
9 VDC line converter
requirements
110 or 220 VAC.
110 or 220 VAC.
for either 110 or 220 VAC.
for either 110 or 220 VAC,
50/60 Hz walloutlet.
50/60 Hz walloutlet.
50/60 Hz wall outlet.
50/60 Hz wall outlet.
input
>10
12
ohms
>10
12
ohms
>10
12
ohms
>10
12
ohms
impedance
instrument drift
< 50 microvolts/oC
< 50 microvolts/oC
< 50 microvolts/oC
< 50 microvolts/oC
input bias current
< ± 1 picoamp at 25oC and < ± 4 picoamps over full operating range
environmental
5 to 45 oC and 5 to 85% relative humidity, non-condensing
requirements
meter dimensions
8.08 x 3.26 x 1.90 inches
case
Dust and Splash Resistant, Chemical Resistant
SPECIFICATIONS
Specifications 125
Summary of Contents for Orion 410Aplus
Page 6: ...2 Orion Benchtop pH ISE Meter Instruction Manual...
Page 84: ...80 Orion Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Page 92: ...Orion Benchtop pH ISE Meter Instruction Manual 88...
Page 104: ...Orion Benchtop pH ISE Meter Instruction Manual 100...
Page 120: ...Orion Benchtop pH ISE Meter Instruction Manual 116...
Page 122: ...Orion Benchtop pH ISE Meter Instruction Manual 118...
Page 126: ...Orion Benchtop pH ISE Meter Instruction Manual 122...