Orion Benchtop pH/ISE Meter Instruction Manual
Special Features
45
Double Known Addition Measurement Procedure
For most double known addition analysis, it is recommended that:
a. Sample volume be 100 mL.
b. The concentration of the standard added be 100 times the
expected sample concentration.
c. The volume of the first addition be 1 mL.
d. The volume of the second addition be 10 mL.
The following table is a sequence of key presses and prompts displayed
when performing a double known addition. (Numbers and electrode ID are
used for illustration purposes only.)
Prompt
Key Press/Action
1. Press
2nd
then
incr tech
.
2. 1-ADDN 2-SUBTR
Press
1
, then
yes
.
3. KNOWN ADDN
The meter verifies your choice.
Wait for the next prompt.
4. 1-SINGLE 2-DOUBLE
Press
2
, then
yes
.
5. DOUBLE
The meter verifies your choice
KNOWN ADDN
then continues.
6. SAMPLE VOLUME?
Enter the volume of the sample.
Press
yes
.
7. TOTAL VOLUME?
Enter the total volume in the beaker.
For example if you have 100 mL of
sample and 10 mL of diluent, enter
110. Press
yes
.
8. STD CONC?
Enter the concentration of the
standard to be added to the sample.
Press
yes
.
9. 1-NA+ IN SAMPLE?
Rinse electrode(s) and place into
sample and press
yes
.
10. MEASURING SAMPLE
When the electrode signal is stable
1-NA+ READY
press
yes
.
YES TO CONTINUE
11. 1ST STD VOLUME?
Enter the volume of standard to be
added to the sample. Press
yes
.
12. ST STD ADDED?
Add standard to sample, then
press
yes
.
Summary of Contents for Orion 410Aplus
Page 6: ...2 Orion Benchtop pH ISE Meter Instruction Manual...
Page 84: ...80 Orion Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Page 92: ...Orion Benchtop pH ISE Meter Instruction Manual 88...
Page 104: ...Orion Benchtop pH ISE Meter Instruction Manual 100...
Page 120: ...Orion Benchtop pH ISE Meter Instruction Manual 116...
Page 122: ...Orion Benchtop pH ISE Meter Instruction Manual 118...
Page 126: ...Orion Benchtop pH ISE Meter Instruction Manual 122...