Orion Benchtop pH/ISE Meter Instruction Manual
Overview of Meter Functions-420Aplus,710Aplus
99
Code
Function
Description
6-2
DATALOG
Sets the timed datalogger interval. The
INTERVAL
maximum interval is Interval 23 hours, 59
minutes, and 59 seconds. The minimum
interval is five (5) seconds. When the code
6-2 is displayed in the lower field, the current
datalogger interval hours are displayed in
the main field indicated by an H in the left
position. Press or to change the setting,
then press
yes
. For the current datalogger
interval minutes: seconds will be displayed in
the main field. Press or to change, then
press
yes
. The default setting is (5) seconds.
6-3
TRANSFER
Transfers data from the datalogger to a
DATA
printer or a computer. When the code 6-3 is
displayed the OFF will flash. To transfer the
data, press or until ON is displayed,
then press
yes
. Once the data has been
transferred, the meter advances to the next
setup option.
NOTE
: Downloading the data does not clear
the memory. To clear memory, see step 6-4 below.
6-4
CLEAR
Clears all data from the datalogger memory.
DATA
When the code 6-4 is displayed, the OFF will
flash. To clear the data, press or until ON
is displayed. The alarm will ring rapidly to
indicate the memory is about to be erased.
Press
yes
.
▼
▼
▼
▼
▼
▼
▼
▼
Summary of Contents for Orion 410Aplus
Page 6: ...2 Orion Benchtop pH ISE Meter Instruction Manual...
Page 84: ...80 Orion Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Page 92: ...Orion Benchtop pH ISE Meter Instruction Manual 88...
Page 104: ...Orion Benchtop pH ISE Meter Instruction Manual 100...
Page 120: ...Orion Benchtop pH ISE Meter Instruction Manual 116...
Page 122: ...Orion Benchtop pH ISE Meter Instruction Manual 118...
Page 126: ...Orion Benchtop pH ISE Meter Instruction Manual 122...