Orion Benchtop pH/ISE Meter Instruction Manual
Special Features
46
Prompt
Key Press/Action
13. MEAS’G SAMP+ STD
Wait for a stable display, then
1-NA+ REA
press
yes.
YES TO CONTINUE
14. 2ND STD VOLUME?
Enter the volume of the second
addition. Press
yes.
15. 2ND STD ADDED
Add the second standard addition to
the sample beaker. Press
yes.
16. MEAS’G SAMP + STD
Wait for a stable display and
1-NA+ READY
press
yes
.
YES TO CONTINUE
17. SLOPE
The meter calculates and displays the
electrode slope for 5 seconds
then continues.
18. DKA RESULT
Meter displays results of analysis in
the main display. Press
yes
.
19. REPEAT TECHNIQUE?
Press
yes
to repeat the technique
without leaving incremental
techniques mode. Repeat steps 6
through 18. Press
no
to exit and
return to measure mode.