Calibration and Measurement of pH
17
B. Orion 410Aplus pH Calibration and Measurement
Autocalibration
Autocalibration is a feature of the Orion 410Aplus Meter that automatically
recognizes the standard buffers, 4.01, 7.00 and 10.01, within a range of
± 0.5 pH units. Simply select the buffer sequence that best fits your
application. Results greater than ± 0.5 pH units from the theoretical value
will trigger an operator assistance code. During calibration, the user waits
for a stable pH reading. Once the electrode is stable, the meter
automatically recognizes and displays the temperature-corrected value for
that buffer. Pressing
yes
enters the value into memory.
Manual Calibration
To calibrate with buffers other than 4.01, 7.00, or 10.01, use the manual
calibration technique. The chosen buffers must be greater than one (1) pH
unit but less than four (4) pH units from the next closest buffer. Manual
calibration is selected through SETUP mode. The calibration sequence is
the same as autocalibration except that the buffer values are scrolled in and
then entered.
NOTE:
For manual calibration use
yes
to accept or
no
to change each digit until the correct value, then press
yes
to accept.
Three Point pH Calibration
Up to a three (3) point calibration can be performed on the Orion 410Aplus
Meter. Simply select the buffer sequence that best fits your application.
Summary of Contents for Orion 410Aplus
Page 6: ...2 Orion Benchtop pH ISE Meter Instruction Manual...
Page 84: ...80 Orion Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Page 92: ...Orion Benchtop pH ISE Meter Instruction Manual 88...
Page 104: ...Orion Benchtop pH ISE Meter Instruction Manual 100...
Page 120: ...Orion Benchtop pH ISE Meter Instruction Manual 116...
Page 122: ...Orion Benchtop pH ISE Meter Instruction Manual 118...
Page 126: ...Orion Benchtop pH ISE Meter Instruction Manual 122...