Troubleshooting
Troubleshooting Guides
6-2
Model 48
i
Instruction Manual
Thermo Electron Corporation
Table 6-2
lists all the alarm messages you may see on the graphics display
and provides recommendations about how to resolve the alarm condition.
Table 6-1.
Troubleshooting - General Guide
Malfunction
Possible Cause
Action
Does not start
(The light on power
switch does not come
on and the pump
motor is not running.)
No power or wrong power
configuration
Check the line to confirm
that power is available and
that it matches the voltage
and frequency configuration
of the instrument.
Main fuse is blown or missing
Unplug the power cord, open
the fuse drawer on the back
panel, and check the fuses
visually or with a multimeter.
Bad switch or wiring connection
Unplug the power cord,
disconnect the switch and
check operation with a
multimeter.
Pressure transducer
does not hold
calibration or is noisy
Pressure transducer defective
Replace pressure transducer.
Run output noisy
Recorder noise
Replace or repair recorder.
Sample CO concentration varying
Run instrument on a span CO
source - if quiet, there is no
malfunction.
Foreign material in optical bench
Clean optical bench.
Analyzer does not
calibrate properly
System leak
Find and repair leak
Pressure or temperature transducer
out of calibration
Recalibrate pressure and
temperature transducer.
Dirty system
Clean cells and flow
components.
Leaky correlation wheel
Replace with a known good
wheel.
Analog test ramp
Faulty recorder
Replace recorder.
D/A calibration off
Recalibrate the D/A with a
DVM known to be in
calibration.
Summary of Contents for 48i
Page 12: ...Preface Thermo Electron Corporation Model 48i Instruction Manual xii ...
Page 22: ...Installation Startup 2 6 Model 48i Instruction Manual Thermo Electron Corporation ...
Page 222: ...Servicing Service Locations 7 40 Model 48i Instruction Manual Thermo Electron Corporation ...
Page 241: ...Thermo Electron Corporation Model 48i Instruction Manual 9 5 Figure 9 3 Bench Mounting ...
Page 242: ...9 6 Model 48i Instruction Manual Thermo Electron Corporation Figure 9 4 EIA Rack Mounting ...