Operation
Instrument Controls Menu
Thermo Electron Corporation
Model 48
i
Instruction Manual
3-71
User Value
The User Value screen allows the user to set the output value for the
corresponding input voltage for the selected table point in the conversion
table, from -999.9 to 999.9. The default table is a two-point table with
point 1: 0.00 V = 000.0 U and point 2: 10.00 V = 10.0 U.
•
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Table Point >
User Value
.
•
Use
and
to move the cursor up and down.
•
Press
to save the new value.
•
Press
to return to the Table Point submenu or
to return
to the Run screen.
Temperature
Compensation
Temperature compensation provides compensation for any changes to the
instrument's output signal due to internal instrument temperature
variations. The effects of internal instrument temperature changes on the
analyzer's subsystems and output have been empirically determined. This
empirical data is used to compensate for any changes in temperature. This
compensation can be used for special applications, or when operating the
instrument outside the recommended temperature range, even though the
Model 48
i
does not require temperature compensation for EPA equivalency.
When temperature compensation is on, the display shows the current
internal instrument temperature (measured by a thermistor on the Interface
board). When temperature compensation is off, the display shows the
factory standard temperature of 25 °C.
XXLO W
CONCENTRATION
XXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARM
x
TABLE POINT 01 USER VAL:
CURRENTLY: 10.800
SET TO: 00.00 ?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Summary of Contents for 48i
Page 12: ...Preface Thermo Electron Corporation Model 48i Instruction Manual xii ...
Page 22: ...Installation Startup 2 6 Model 48i Instruction Manual Thermo Electron Corporation ...
Page 222: ...Servicing Service Locations 7 40 Model 48i Instruction Manual Thermo Electron Corporation ...
Page 241: ...Thermo Electron Corporation Model 48i Instruction Manual 9 5 Figure 9 3 Bench Mounting ...
Page 242: ...9 6 Model 48i Instruction Manual Thermo Electron Corporation Figure 9 4 EIA Rack Mounting ...