Operation
Service Menu
Thermo Electron Corporation
Model 48
i
Instruction Manual
3-109
Initial S/R Ratio
The Initial S/R Ratio screen displays both the initial S/R ratio and the
current S/R ratio. The initial S/R ratio is determined at the factory, and is
used to correct for the slight variations found from one correlation wheel to
another. The only time the initial S/R ratio should be changed is when the
correlation wheel is replaced or sample/reference is between 1.14 and 1.18.
•
In the Main Menu, choose Service >
Initial S/R Ratio
.
•
Use the
and
to move and change the value.
•
Press
to save set to value as actual value.
•
Press
to return to the Service menu or
to return to the
Run screen.
XXLO W
CONCENTRATION
XXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARM
X
RESTORE DEFAULT CAL:
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
¨
TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM
WARNING
This adjustment should only be performed by an instrument
service technician.