Operation
Alarms Menu
Thermo Electron Corporation
Model 48
i
Instruction Manual
3-97
Motor Speed
The Motor Speed screen (read only) displays the current motor speed. A
reading other than 100.0% indicates a problem with either the motor or the
power supply.
•
In the Main Menu, choose Alarms >
Motor Speed
.
•
Press
to return to the Alarms menu or
to return to the
Run screen.
Zero and Span Check
The Zero Span Check screen allows the user to view the status of the most
recent zero check and set the maximum zero check offset. The zero and
span check screens are visible only if the zero/span check option is enabled
and function the same way.
•
In the Main Menu, choose Diagnostics > Voltages >
Zero
or
Span
Check
.
•
Use the
and
to move the cursor up and down.
•
Press
to select a choice.
•
Press
to return to the Alarms menu or
to return to the
Run screen.
XXLO W
CONCENTRATION
XXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARM
x
MOTOR SPEED:
ACTUAL 100.0 %
RANGE AVG DIAGS ALARM