MODBUS Protocol
MODBUS Commands Supported
C-8
48
i
Instruction Manual
Thermo Electron Corporation
MODBUS Commands
Supported
The following Tables 1–3 lists the MODBUS commands supported for the
Model 48
i
.
Response
Field Name
(Hex)
Function
05
Output Address Hi
00
Output Address Lo
05
Output Value Hi
FF
Output Value Lo
00
Table C-1.
Read Registers for 48
i
Register Number
Variable
40001&40002
CO
40003&40004
LO CO
40005&40006
HI CO
40007&40008
RANGE STATUS
40009&40010
S/R
40011&40012
LO S/R
40013&40014
HI S/R
40015&40016
INT TEMP
40017&40018
BENCH TEMP
40019&40020
AUX TEMP
40021&40022
PERM OVN GAS
40023&40024
PERM OVN HTR
40025&40026
BENCH PRES
40027&40028
SAMPLE FLOW
40029&40030
INTENSITY
40031&40032
MOTOR SPEED
40033&40034
ANALOG IN 1
40035&40036
ANALOG IN 2
40037&40038
ANALOG IN 3
40039&40040
ANALOG IN 4
40041&40042
ANALOG IN 5
40043&40044
ANALOG IN 6