Introduction
Specifications
Thermo Electron Corporation
48
i
Instruction Manual
1-3
Figure 1-1.
Model 48
i
Flow Schematic
Specifications
Table 1-1.
Model 48
i
Specifications
Preset ranges
0-1, 2, 5, 10, 20, 50, 100, 200, 500, 1000, 2000, 5000, 10000 (ppm
or mg/m
3
)
Custom ranges
0-1 to 10000 (ppm or mg/m
3
)
Zero roise
0.02 ppm RMS (30 second averaging time)
Lower detectable limit
0.04 ppm
Zero drift (24 hour)
< 0.1 ppm
Span drift
±
1% full-scale
Response time
60 seconds (30 second averaging time)
Linearity
±
1% of full-scale
≤
1000 ppm
±
2.5% of full-scale
>
1000 ppm
Sample flow rate
1.0 LPM
Operating temperature
20
–
30 °C (may be safely operated over the range of 0
–
45 °C)
*