Calibration
HI and LO Multi-Point Calibration
Thermo Electron Corporation
Model 48
i
Instruction Manual
4-9
3. From the Main Menu select Service Mode > Lo Multi-Point Cal >
Cal
Point 1
.
4. Use
to move the cursor and
to increment or
decrement the digit values until they match the concentration being
introduced to the instrument. Press
.
The instrument will perform a series of calculations and save the new
parameters.
5. Press
to back up a step in the Service Mode menu.
6. Repeat Step 1 for a 50% concentration of the upper range.
7. Select Cal Point 2.
8. Use
to move the cursor and
to increment or
decrement the digit values until they match the concentration being
introduced to the instrument. Press
.
The instrument will perform a series of calculations and save the new
parameters.
9. Press
to back up a step in the Service Mode menu.
10. Repeat Step 1 for a 20% concentration of the upper range.
11. Select Cal Point 3.
12. Use
to move the cursor and
13.
to increment or decrement the digit values until they match the
concentration being introduced to the instrument. Press
.
The instrument will perform a series of calculations and save the new
parameters.
Summary of Contents for 48i
Page 12: ...Preface Thermo Electron Corporation Model 48i Instruction Manual xii ...
Page 22: ...Installation Startup 2 6 Model 48i Instruction Manual Thermo Electron Corporation ...
Page 222: ...Servicing Service Locations 7 40 Model 48i Instruction Manual Thermo Electron Corporation ...
Page 241: ...Thermo Electron Corporation Model 48i Instruction Manual 9 5 Figure 9 3 Bench Mounting ...
Page 242: ...9 6 Model 48i Instruction Manual Thermo Electron Corporation Figure 9 4 EIA Rack Mounting ...