Servicing
Front Panel Board Replacement
7-36
Model 48
i
Instruction Manual
Thermo Electron Corporation
Figure 7-17.
Replacing the Measurement Interface Board
Front Panel Board
Replacement
Use the following procedure to replace the front panel board (
Figure 7-18
).
Equipment Required:
•
Front panel board
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Remove the three ribbon cables and the two-wire connector from the
front panel board.
3. Pop off the board from the two top mounting studs and remove the
board by lifting it up and off the slotted bottom support.
4. Replace the front panel board by following previous steps in reverse.
Measurement Interface Board
Mounting Studs (4)
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.