Operation
Diagnostics Menu
Thermo Electron Corporation
Model 48
i
Instruction Manual
3-81
Sample/Reference Ratio
The Sample/Reference Ratio screen (read only) displays the ratio of the
intensities of the light source through the sample (CO) side and reference
(N2) side of the correlation wheel. Normally, when zero air is being
sampled, the sample/reference ratio is between 1.14 and 1.18. A ratio
outside may indicate that the correlation wheel is dirty or leaking gas.
•
In the Main Menu, choose Diagnostics >
Sample/Ref Ratio
.
•
Press
to return to the Diagnostics menu or
to return to
the Run screen.
AGC Intensity
The AGC Intensity screen (read only) displays the intensity (in Hertz) of
the reference channel Automatic Gain Control (AGC) circuit. The AGC
circuit optimizes the noise and resolution levels of the Model 48
i
. The AGC
intensity reading should be about 250,000 Hertz.
•
In the Main Menu, choose Diagnostics >
AGC Intensity
.
•
Press
to return to the Diagnostics menu or
to return to
the Run screen.
XXLO W
CONCENTRATION
XXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARM
x
SAMPLE REFERENCE RATIO:
S/R 1.1611
RANGE AVG DIAGS ALARM
Summary of Contents for 48i
Page 12: ...Preface Thermo Electron Corporation Model 48i Instruction Manual xii ...
Page 22: ...Installation Startup 2 6 Model 48i Instruction Manual Thermo Electron Corporation ...
Page 222: ...Servicing Service Locations 7 40 Model 48i Instruction Manual Thermo Electron Corporation ...
Page 241: ...Thermo Electron Corporation Model 48i Instruction Manual 9 5 Figure 9 3 Bench Mounting ...
Page 242: ...9 6 Model 48i Instruction Manual Thermo Electron Corporation Figure 9 4 EIA Rack Mounting ...