Alarms
Thermo Electron Corporation
Model 48
i
Instruction Manual
B-13
These commands set the CO concentration alarm minimum and maximum
values to
value,
where
value
is a floating-point representation of the
concentration alarm limits. Values must be in the units that are currently set
for use. The example below sets the CO concentration alarm maximum
value to 215.
Send:
set alarm conc max 215
Receive:
set alarm conc max 215 ok
alarm internal temp min
alarm internal temp max
These commands report the internal temperature alarm minimum and
maximum value current settings. The example below reports that the
internal temperature alarm minimum value is 15.0 °C.
Send:
internal temp alarm min
Receive:
internal temp alarm min 15.0 deg C
set internal temp alarm min
value
set internal temp alarm max
value
These commands set the internal temperature alarm minimum and
maximum values to
value
, where
value
is a floating-point number
representing internal temperature alarm limits in degrees C. The example
below sets the internal temperature alarm maximum value to 45.0 °C.
Send:
set internal temp alarm max 45
Receive:
set internal temp alarm max 45 ok
alarm motor speed min
alarm motor speed max
These commands report the motor speed alarm minimum and maximum
value current settings. The example below reports that the motor speed
alarm minimum value is 20 minutes.
Send:
alarm motor speed min
Receive:
alarm motor speed min 20
set alarm motor speed min
value
set alarm motor speed max
value
These commands set the motor speed alarm minimum and maximum
values to
value
, where
value
is a floating-point number representing motor
speed alarm limits in minutes. The example below sets the motor speed
alarm maximum value to 20 minutes.
Send:
set alarm motor speed max 20
Receive:
set alarm motor speed max 20 ok
alarm pressure min
Summary of Contents for 48i
Page 12: ...Preface Thermo Electron Corporation Model 48i Instruction Manual xii ...
Page 22: ...Installation Startup 2 6 Model 48i Instruction Manual Thermo Electron Corporation ...
Page 222: ...Servicing Service Locations 7 40 Model 48i Instruction Manual Thermo Electron Corporation ...
Page 241: ...Thermo Electron Corporation Model 48i Instruction Manual 9 5 Figure 9 3 Bench Mounting ...
Page 242: ...9 6 Model 48i Instruction Manual Thermo Electron Corporation Figure 9 4 EIA Rack Mounting ...