Operation
Service Menu
3-104
Model 48
i
Instruction Manual
Thermo Electron Corporation
Calibrate Pressure Zero
The Calibrate Pressure Zero screen calibrates the pressure sensor at zero
pressure.
Note
A vacuum pump must be connected to the pressure sensor before
performing the zero calibration.
•
In the Main Menu, choose Service > Pressure Calibration >
Zero
.
•
Press
to save the current pressure reading as the zero reading.
•
Press
to return to the Pressure Calibration menu or
to
return to the Run screen.
Calibrate Pressure Span
The Calibrate Pressure Span screen allows the user to view and set the
pressure sensor calibration span point.
XXLO W
CONCENTRATION
XXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARM
X
PRESSURE SENSOR CAL:
>ZERO 200
SPAN 1.2200
SET DEFAULTS
RANGE AVG DIAGS ALARM
XXLO W
CONCENTRATION
XXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARM
X
CALIBRATE PRESSURE ZERO:
CURRENTLY: 0.0 mmHg
SET TO: 0.0 mmHg ?
CONNECT VACUUM PUMP AND
SAVE ZERO PRESSURE
RANGE AVG DIAGS ALARM
Summary of Contents for 48i
Page 12: ...Preface Thermo Electron Corporation Model 48i Instruction Manual xii ...
Page 22: ...Installation Startup 2 6 Model 48i Instruction Manual Thermo Electron Corporation ...
Page 222: ...Servicing Service Locations 7 40 Model 48i Instruction Manual Thermo Electron Corporation ...
Page 241: ...Thermo Electron Corporation Model 48i Instruction Manual 9 5 Figure 9 3 Bench Mounting ...
Page 242: ...9 6 Model 48i Instruction Manual Thermo Electron Corporation Figure 9 4 EIA Rack Mounting ...