Operation
Service Menu
Thermo Electron Corporation
Model 48
i
Instruction Manual
3-105
Note
The plumbing going to the pressure sensor should be disconnected so
the sensor is reading ambient pressure before performing the span
calibration. The operator should use an independent barometer to measure
the ambient pressure and enter the value on this screen before calibrating.
•
In the Main Menu, choose Service > Pressure Calibration >
Span
.
•
Use
,
,
and
to move and change the value.
•
Press
to save set to value as actual value.
•
Press
to return to the Pressure Calibration menu or
to
return to the Run screen.
Restore Default Pressure
Calibration
The Restore Default Pressure Calibration screen allows the user to reset the
pressure calibration configuration values to factory defaults.
•
In the Main Menu, choose Service > Pressure Calibration >
Set
Defaults
.
•
Press
to warn user and enable restore with
.
•
Use
to overwrite pressure sensor calibration parameters with
factory default values when pressed after
.
•
Press
to return to the Pressure Calibration menu or
to
return to the Run screen.
XXLO W
CONCENTRATION
XXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARM
X
CALIBRATE PRESSURE SPAN:
CURRENTLY: 750.0 mmHg
SET TO: 760.0 mmHg ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Summary of Contents for 48i
Page 12: ...Preface Thermo Electron Corporation Model 48i Instruction Manual xii ...
Page 22: ...Installation Startup 2 6 Model 48i Instruction Manual Thermo Electron Corporation ...
Page 222: ...Servicing Service Locations 7 40 Model 48i Instruction Manual Thermo Electron Corporation ...
Page 241: ...Thermo Electron Corporation Model 48i Instruction Manual 9 5 Figure 9 3 Bench Mounting ...
Page 242: ...9 6 Model 48i Instruction Manual Thermo Electron Corporation Figure 9 4 EIA Rack Mounting ...