Operation
Instrument Controls Menu
3-52
Model 48
i
Instruction Manual
Thermo Electron Corporation
Gateway
The Gateway screen is used to edit the gateway address.
•
In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings
> Gateway
.
•
Use
,
,
and
to move and change the value of
the gateway address.
•
Press
to save the new address.
•
Press
to return to the TCP/IP Settings menu or
to return
to the Run screen.
Host Name
The host name screen is used to edit the host name. When DHCP is
enabled, this name is reported to the DHCP server.
X W
CONCENTRATION
W
CO 15.80 PPM
SAMPLE 12:34 ALARM
x
NETMASK:
CURRENT: 255.255.255.000
SET TO: 255.255.255.00
0
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
X W
CONCENTRATION
W
CO 15.80 PPM
SAMPLE 12:34 ALARM
x
GATEWAY:
CURRENT: 000.000.000.000
SET TO: 192.168.001.00
1
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM