Operation
Service Menu
3-114
Model 48
i
Instruction Manual
Thermo Electron Corporation
Temperature Calibration
The Termperature calibration screen allows the user to view and set the
ambient temperature sensor calibration.The temperature calibration is
visible only when the instrument is in service mode. For more information
on the service mode, see “Service Mode” earlier in the chapter.
•
In the Main Menu, choose Service >
Temperature Calibration
.
•
Use
,
,
and
to move and change the value.
•
Press
to save set to value as actual value.
•
Press
to return to the Service menu or
to return to the
Run screen.
Analog Output Calibration
The Analog Output Calibation menu is a selection of analog output to
calibrate, and allows the user to select the calibration action zero or span.
The analog output calibration is visible only when the instrument is in
service mode. For more information on the service mode, see “Service
Mode” earlier in the chapter.
WARNING
This adjustment should only be performed by an instrument
service technician.
XXLO W
CONCENTRATION
XXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARM
X
CALIBRATE AMBIENT TEMP:
ENTER TEMP TO CAL TO
SET TO: 25.0
o
C
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Summary of Contents for 48i
Page 12: ...Preface Thermo Electron Corporation Model 48i Instruction Manual xii ...
Page 22: ...Installation Startup 2 6 Model 48i Instruction Manual Thermo Electron Corporation ...
Page 222: ...Servicing Service Locations 7 40 Model 48i Instruction Manual Thermo Electron Corporation ...
Page 241: ...Thermo Electron Corporation Model 48i Instruction Manual 9 5 Figure 9 3 Bench Mounting ...
Page 242: ...9 6 Model 48i Instruction Manual Thermo Electron Corporation Figure 9 4 EIA Rack Mounting ...