Preamp Board Calibration
The Pre-amp Board Calibration screen is used to adjust the preamp board
calibration parameters.
•
In the Main Menu, choose Service >
Preamp Calibration
.
•
Use
and
to increment or decrement the numeric value.
•
Press
to save the new counts to hardware.
•
Press
to return to the Service menu or
to return to the
Run screen.
XXLO W
CONCENTRATION
XXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARM
X
COEFFICIENTS:
1: 1.000000
2: 1.000000
3: 1.000000
RECALCULATE?
RANGE AVG DIAGS ALARM
WARNING
This adjustment should only be performed by an instrument
service technician.
XXLO W
CONCENTRATION
XXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARM
x
PREAMP BOARD CAL:
SAMPLE: 127807
REFERENCE: 148256
SET TO: 128 ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM