Dionex ICS-1100 Ion Chromatography System
62
Doc. 065289-03 10/12
To troubleshoot:
1. Retry the calibration.
2. If calibration still fails, the vacuum degas assembly may need to be
replaced. Contact Thermo Fisher Scientific for assistance.
This error indicates a problem in the Dionex ICS-1100 electronics.
To troubleshoot:
Contact Thermo Fisher Scientific for assistance. The Dionex ICS-1100
electronics components cannot be serviced by the user.
The leak sensor is installed in the drip tray at the bottom of the component
panel (see
). If liquid accumulates in the tray, the sensor signals the
problem and this error message appears.
To troubleshoot:
1. Locate the source of the leak by visually inspecting the tubing, fittings,
and components on the component panel. Refer to
2. Tighten fittings or replace tubing and fittings as required. Refer to
for detailed troubleshooting of various types of leaks.
3. After fixing the leak, dry the drip tray thoroughly to prevent the leak
sensor from triggering additional error messages.
If the injection valve fails to switch position within 1 second of being toggled,
the Dionex ICS-1100 Moduleware reports an error to Chromeleon and this
error message appears.
HARDWARE NOT PRESENT
LEAK SENSOR WET
LOAD/INJECT VALVE ERROR
Summary of Contents for Dionex ICS-1100
Page 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Page 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Page 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Page 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Page 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Page 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Page 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Page 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...