4 • Troubleshooting
Doc. 065289-03 10/12
75
•
Incorrect or contaminated eluent
Remake the eluent. If the system is not operating in the RFIC-ER mode, refer
to the instructions in the column manual. The manual is included on the
Thermo Scientific Reference Library DVD (P/N 053891).
If this is an RFIC-ER system, refer to the instructions in the manual provided
in your RFIC-ER Installation Kit.
Install or replace the end-line filter on the end of the deionized water line (see
•
Malfunctioning injection valve or auxiliary valve
1. Rebuild the valve (see
2. If the valve leak persists, contact Thermo Fisher Scientific for assistance.
4.10 Nonreproducible Peak Height or Retention Time
•
Column overloading
Dilute the sample (see
•
Liquid leaks
Locate and eliminate the leaks (see
•
Incomplete or imprecise filling of the sample loop
1. Fill the sample loop until excess sample exits the waste line.
2. Inspect the 1-cc syringe (P/N 016388) and replace if damaged.
•
Pump not primed properly
4.11 Abnormal Retention Time or Selectivity
•
Incorrect or contaminated eluent
Remake the eluent, using concentrated eluent and ASTM filtered, Type I (18-
megohm) deionized water.
Summary of Contents for Dionex ICS-1100
Page 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Page 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Page 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Page 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Page 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Page 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Page 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Page 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...