5 • Service
Doc. 065289-03 10/12
101
Reinstalling the Head and Piston
Thermo Fisher Scientific recommends reinstalling the head and piston as a single
assembly, so that the piston centers itself onto the magnetic follower.
1. Hold the assembled spacer and guide with the drain tubes aligned vertically
and press the spacer into the head until it is flush with the indented surface of
the head.
2. Insert the piston so that 1/4 inch of the shaft is exposed. This ensures that the
magnet in the follower picks up the piston. (The follower is the cylinder that
holds the piston in place as it moves in and out of the pump head assembly.)
3. Reinstall the head and piston assembly; use a wrench to tighten the nuts
evenly (12 in-lb torque).
Completing the Procedure
1. Reconnect all liquid lines to the pump head.
2. Turn on the main power switch.
3. Open the eluent valve.
5.8
Replacing a Pump Piston
Continued leaking of the piston seal after installation of a new seal (assuming
the pump head is tight) indicates a dirty, scratched, or broken piston.
Follow the instructions in
to install a new piston (P/N 052840) and
piston seal (P/N 055870).
Summary of Contents for Dionex ICS-1100
Page 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Page 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Page 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Page 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Page 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Page 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Page 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Page 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...