Dionex ICS-1100 Ion Chromatography System
50
Doc. 065289-03 10/12
3.10.1 Loading Samples with a Syringe
1. Make sure the injection port on the Dionex ICS-1100 front door (see
) is connected to sample port
S (5)
on the injection valve
2. Fill the 1 cc syringe (P/N 016388) provided in the Dionex ICS-1100
Ship Kit (P/N 057905) with a calibration standard or sample.
3. Insert the syringe into the injection port on the Dionex ICS-1100 front
door (see
4. Verify that the injection valve is in the Load position.
5. Overfill the sample loop with several sample loop volumes. Excess
sample will exit through the injection valve waste line.
6. Leave the syringe in the port.
7. Switch the injection valve to the Inject position (see
Figure 3-7. Injection Valve Connections