5 • Service
Doc. 065289-03 10/12
93
4. Unscrew the black locking ring
on the front of the valve (see
) and remove the
ring.
5. Grasp the front of the valve pod
and pull firmly to remove it
from the Dionex ICS-1100.
Figure 5-5. Unscrew the Valve Locking Ring
6. Check that the new pod (6-port, P/N 061947; 10-port, P/N 061948) has the
correct number of ports for the valve being serviced.
7. Align the slots in the new pod with the runner in the valve holder on the
Dionex ICS-1100 (see
). Valve pods are keyed to fit only one way
(one slot is narrower than the other). Verify that the slots are aligned with their
matching runners.
8. Also verify that the two splines on the pod align with the matching splines
). If necessary, twist the end of the pod
to adjust the position of the splines.
Figure 5-6. Valve Pod and Pod Holder
Black
Locking Ring
Slot (larger)
Splines
Runner (larger)
Valve Pod