3 • Operation and Maintenance
Doc. 065289-03 10/12
55
c.
Click
Next >
and select the
in a new sequence via Sequence
Wizard
option.
d. Click
Next >
to go to the Sequence Wizard.
e.
Complete the steps in the Sequence Wizard, adding the desired
number of samples and standards to the list. For help, click the
Help
button on the Sequence Wizard page.
After you click
Finish
, a sequence is created and a program
appropriate for the selected application is copied to the sequence. If
you are using Chromeleon, a quantification method is also copied to
the sequence
.
5. Load the sequence and start batch processing:
a.
On the
Sequence Control
panel, click
Load Sequence
.
b. Select the sequence created in
and click
Open
.
c.
Click
Start Batch
.
3.12 Maintenance
This section describes routine maintenance procedures that users may perform.
All other maintenance procedures must be performed by Thermo Fisher Scientific
personnel.
As Needed
•
Regularly check the eluent reservoir and refill when needed.
•
Make fresh eluent as needed.
•
RFIC-ER systems only: The Chromeleon RFIC-ER Wellness feature tracks
the usage of the RFIC-ER eluent and columns. An Audit Trail message will
inform you when an item is approaching the end of its useful life span. You
can also check the life expectancy of these components on the RFIC-ER
Wellness panel.
Daily
•
Check the Dionex ICS-1100 component panel (see
spills. Wipe up spills. Isolate and repair leaks (see
). Rinse off any
dried eluent with deionized water.
Summary of Contents for Dionex ICS-1100
Page 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Page 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Page 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Page 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Page 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Page 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Page 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Page 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...