Dionex ICS-1100 Ion Chromatography System
80
Doc. 065289-03 10/12
4. Under
Degas
,
verify that the
Always Off
option is
not
selected. If it is,
select one of the following settings instead:
•
Cycle
; also select
On
: 30 seconds and
Off
: 10 minutes
•
Monitor
5. If the
Degas
settings are correct, but the degas assembly still does not run,
test the assembly by selecting the
Always On
option. The pump should
turn on immediately. If it does not, the vacuum degas assembly may need
to be replaced. Contact Thermo Fisher Scientific for assistance.
6. If the degas pump runs when you select
Always On
, but does not run
when
Monitor
is selected, recalibrate the degas assembly (see
Select the Degas, Always On option only when testing the vacuum degas
assembly. Do not select the Always On option for routine use.