Dionex ICS-1100 Ion Chromatography System
32
Doc. 065289-03 10/12
2.4.5
Auxiliary Valve (Optional)
The auxiliary valve is a high-pressure Rheodyne valve. The electrically-
activated, 2-position PEEK valve is offered in both 6-port and 10-port
models (6-Port Valve Kit, P/N 069472; 10-Port Valve Kit, P/N 069473).
The auxiliary valve enables a variety of sample preparation activities,
including:
•
Online filtration
•
Matrix elimination (for example, the removal of high backgrounds of
chloride or organic material)
•
Concentrator-based techniques
•
Conditional injections (large loop/small loop applications where the
data system monitors sample concentration and reinjects the sample,
using the smaller loop, if the concentration is too high)
•
AutoNeutralization
™
•
Matrix diversion prior to MS (mass spectrometry) detection
For more information, refer to
Installing the ICS-1100/ICS-1600/ICS-
2100 Auxiliary Valve
(Document No. 065288). The manual is provided in
the Auxiliary Valve Kit.
Figure 2-14. Injection Valve Plumbing
Summary of Contents for Dionex ICS-1100
Page 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Page 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Page 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Page 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Page 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Page 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Page 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Page 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...