Dionex ICS-1100 Ion Chromatography System
116
Doc. 065289-03 10/12
5.17 Priming the Pump with Isopropyl Alcohol
NOTE Prime the pump heads with isopropyl alcohol only if the
two standard priming procedures described in
are unsuccessful.
1. Connect a 10 mL syringe
(P/N 079803) filled with
isopropyl alcohol (IPA) to
the port in the primary
pump head (see
2. Open the waste valve on
the secondary pump head
(see
turning the knob one-
quarter to one-half turn
counterclockwise.
Figure 5-25. Priming the Pump Heads with IPA
3. In the
Pump Settings
Pump On
.
4. Open the priming valve on the primary pump head by turning it one-quarter to
one-half turn counterclockwise.
5. Use the syringe to slowly push alcohol through the pump.
NOTE Be careful not to push any air trapped in the syringe
through the pump. Check the waste line from the
secondary pump head to verify that there are no air
bubbles.
6. Close the priming valve.
Do not overtighten.
Disconnect the syringe from
the priming valve.
7. Let the pump run for several minutes, to purge alcohol from the pump heads.
Then, click the
Prime
button in the
Pump Settings
window to flush the heads
with the desired eluent.
10 mL Syringe
(P/N 079803)
Priming Valve
(open)
Port
Waste Valve
(open)
Primary
Pump Head
Secondary
Pump Head
Summary of Contents for Dionex ICS-1100
Page 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Page 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Page 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Page 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Page 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Page 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Page 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Page 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...