2 • Features
Doc. 065289-03 10/12
31
Eluent flows through either the Load or Inject path, depending on the
valve position.
•
In the Load position, sample is loaded into the sample loop, where it
is held until injection. Eluent flows from the pump, through the valve,
and to the column, bypassing the sample loop. Sample flows from the
syringe or automated sampler line (if installed), through the valve,
and into the sample loop. Excess sample flows out to waste.
•
In the Inject position, sample is swept to the column for analysis.
Eluent flows from the pump, through the sample loop, and on to the
column, carrying the contents of the sample loop with it.
describes how to inject samples.
shows the injection valve connections. The injection valve is
plumbed at the factory with all tubing and fittings for connection to the
pump, injection port, column, and waste. A 25-
L PEEK
™
sample loop
(P/N 042857) is installed between ports L (1) and L (4). Thermo Fisher
Scientific offers sample loops in various sizes. If necessary, the pre-
installed 25-
L loop can be replaced with a loop that has a different
sample injection volume.
Figure 2-13. Injection Valve Flow Schematics
To Waste
Sample In
To Column
From Pump
Sample
Loop
LOAD POSITION
To Waste
Sample In
From Pump
Sample
Loop
INJECT POSITION
To Column
Summary of Contents for Dionex ICS-1100
Page 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Page 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Page 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Page 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Page 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Page 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Page 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Page 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...